10A NCAC 15 .0207 ADDITIONAL REQUIREMENTS TO PROVIDE SERVICES
(a) A person applying for registration to perform Class II or Class IX services for diagnostic radiation output measurements and equipment surveys, Class V area radiation surveys and shielding evaluations for radiographic and fluoroscopic X-ray machines, or Class VII therapeutic area radiation survey or shielding verification services pursuant to Rule .0205 of this Section shall meet the following additional requirements:
(1) have radiation survey and radiation measurement equipment capable of measuring the radiation energies corresponding to the services requested for authorization;
(2) ensure that the equipment in Subparagraph (a)(1) of this Rule is calibrated annually when a frequency is not recommended by the manufacturer. A frequency other than annually shall be accepted if stated in the manufacturer's manual.
(3) submit the following for agency review prior to registration:
(A) a description of the procedures that will be used in performing area radiation surveys, including a list of all guides and references provided to the employed;
(B) a copy of all forms, reports, and documents that will be supplied in conjunction with the provided services;
(C) samples of surveys for each modality requested for registration;
(D) samples of reports of diagnostic radiation output measurements for each modality requested for registration; and
(E) samples of calibration reports for each therapeutic and kV imaging modality requested for registration.
(b) A person applying for registration to perform Class VI equipment calibrations shall meet the following requirements:
(1) ensure such calibrations are current at the time that the service is provided and traceable to the National Institute of Standards and Technology;
(2) license or register radiation sources used for such calibration as required by the rules in this Chapter;
(3) label the equipment to indicate the date of calibration; and
(4) maintain records of the calibration.
(c) A person applying for registration to perform Class III shielding designs for X-ray machines, Class IV shielding designs for fluoroscopic X-ray machines, and Class VII therapeutic shielding design verification services shall meet the following additional requirements:
(1) submit examples of the shielding design which will be provided to registrants;
(2) submit any technical guides, methodology, occupancy factor rationales, and workload estimation rationales that will be used; and
(3) ensure that the shielding design verification services provided to registrants meet the requirements in this Chapter.
History Note: Authority G.S. 104E‑7;
Eff. February 1, 1980;
Amended Eff. June 1, 1993; June 1, 1989;
Transferred and Recodified from 15A NCAC 11 .0207 Eff. February 1, 2015.
Readopted Eff. October 1, 2025;
Amended Eff. October 1, 2026.