10A NCAC 15 .0207         ADDITIONAL REQUIREMENTS TO PROVIDE SERVICES

(a)  A person applying for registration to perform Class II or Class IX services for diagnostic radiation output measurements and equipment surveys, Class V area radiation surveys and shielding evaluations for radiographic and fluoroscopic X-ray machines, or Class VII therapeutic area radiation survey or shielding verification services pursuant to Rule .0205 of this Section shall meet the following additional requirements:

(1)           have radiation survey and radiation measurement equipment capable of measuring the radiation energies corresponding to the services requested for authorization;

(2)           ensure that the equipment in Subparagraph (a)(1) of this Rule is calibrated annually when a frequency is not recommended by the manufacturer. A frequency other than annually shall be accepted if stated in the manufacturer's manual.

(3)           submit the following for agency review prior to registration:

(A)          a description of the procedures that will be used in performing area radiation surveys, including a list of all guides and references provided to the employed;

(B)          a copy of all forms, reports, and documents that will be supplied in conjunction with the provided services;

(C)          samples of surveys for each modality requested for registration;

(D)          samples of reports of diagnostic radiation output measurements for each modality requested for registration; and

(E)           samples of calibration reports for each therapeutic and kV imaging modality requested for registration.

(b)  A person applying for registration to perform Class VI equipment calibrations shall meet the following requirements:

(1)           ensure such calibrations are current at the time that the service is provided and traceable to the National Institute of Standards and Technology;

(2)           license or register radiation sources used for such calibration as required by the rules in this Chapter;

(3)           label the equipment to indicate the date of calibration; and

(4)           maintain records of the calibration.

(c)  A person applying for registration to perform Class III shielding designs for X-ray machines, Class IV shielding designs for fluoroscopic X-ray machines, and Class VII therapeutic shielding design verification services shall meet the following additional requirements:

(1)           submit examples of the shielding design which will be provided to registrants;

(2)           submit any technical guides, methodology, occupancy factor rationales, and workload estimation rationales that will be used; and

(3)           ensure that the shielding design verification services provided to registrants meet the requirements in this Chapter.

 

History Note:        Authority G.S. 104E‑7;

Eff. February 1, 1980;

Amended Eff. June 1, 1993; June 1, 1989;

Transferred and Recodified from 15A NCAC 11 .0207 Eff. February 1, 2015.

Readopted Eff. October 1, 2025;

Amended Eff. October 1, 2026.